Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-63

Chapter 13

Occlusion
pressure at 100
ms (P0.1)
maneuver

Procedure: No instructions to the patient are required. Available in VCV and
PCV. Not available in NPPV.

During the maneuver: Settings do not change during the P0.1 maneuver.
The ventilator briefly occludes the patient circuit and measures occlusion
pressure in the first 100 ms of four preselected patient initiated breaths. If
you press Stop during a P0.1 maneuver, results are displayed for the
maneuver breaths that did occur.

Results display: P0.1 is displayed in cmH

2

O, and is time-stamped based on

ventilator time and date. Measured pressures from –100 to +200 cmH

2

O

are displayed for the P0.1 maneuver: If measured pressure exceeds these
limits, only a message (no data) is displayed.

Range: -100 to +200 cmH

2

O

Resolution: 1 cmH

2

O

Accuracy: ± (0.5 cmH

2

O) ± (10%) of true value

Parameter

Description

Table 13-9: Respiratory Mechanics Maneuver Summary and Data Specifications