Appendix A
RS-232 Communications Protocol
REF 1057983 A
Respironics V200 Ventilator Operator’s Manual
Appendix A-3
VCV E-Cycle Type
%Flow
◆
N/A
%FLOW
◆
AUTO
◆◆
N/A
VCV %Flow Expiratory
Cycle Setting
25
◆◆◆◆
1
10 - 80
%
VCV Rise Time
0.1
◆◆◆
0.1
0.1 - 0.9
Sec
VCV Oxygen
Concentration Setting
21
◆◆◆◆
1
21 - 100
%
VCV Plateau Setting
0.0
◆◆◆
0.1
0.0 - 2.0
Sec
VCV Apnea Rate
Setting
12
◆◆◆◆
1
1 - 80
BPM
VCV High Inspiratory
Pressure Alarm Limit
Setting
35
◆◆◆◆
1
10 - 105
cmH
2
O
(hPa)
VCV Low Inspiratory
Pressure Alarm Limit
Setting
3
◆◆◆◆◆
1
3 - 105
cmH
2
O
(hPa)
VCV Low PEEP Alarm
Limit Setting
0
◆◆◆◆◆
1
0 - 35
cmH
2
O
(hPa)
VCV Low Mandatory
Tidal Volume Alarm
Limit Setting
0
◆◆◆◆◆
1
0 - 2500
mL
VCV Low Spontaneous
Tidal Volume Alarm
Limit Setting
0
◆◆◆◆◆
1
0 - 2500
mL
VCV High Respiratory
Rate Alarm Limit
Setting
150
◆◆◆
1
0 - 150
BPM
VCV Low Minute
Volume Alarm Limit
Setting
1.00
◆◆
0.01 for
0.00 to 9.99
0.1 for 10.0
to 60.0
0.00 - 60.0
L
VCV Apnea Interval
Alarm Limit Setting
20
◆◆◆◆
1
10 - 60
sec
PCV Mode Setting
A/C
◆◆◆
N/A
A/C
◆◆◆
SIMV
◆◆
CPAP
◆◆
N/A
PCV Patient Type
ADULT
◆
N/A
ADULT
◆
PED
◆◆◆
NEO
◆◆◆
PCV Respiratory Rate
Setting
12
◆◆◆◆
1
1 - 150
BPM
PCV Pressure Setting
20
◆◆◆◆
1
5 - 100
cmH
2
O
(hPa)
Ventilation Report (Continued)
Description
Example
Resolution
Range
Units
Comments
Table A-1: Ventilation Report (Sheet 2 of 10)