Warnings, Cautions, and Notes
•
If the EXP HOLD key is held continuously, and the expiratory hold 
maneuver exceeds 5 seconds, the ventilator automatically 
terminates the expiratory hold maneuver and begins a new 
inspiratory period.
•
If Auto PEEP as calculated in Equation 1: Auto-PEEP = Expiratory 
Pause Pressure – End Expiratory Pressure, is negative, Auto-PEEP 
will be displayed as “—.”
•
All components of the patient circuit must not have leaks in order 
to pass SST.
•
If time is found to be incorrect more than once in the 
preoperational procedure, an internal battery may have to be 
replaced. Contact qualified service personnel or call Respironics 
Customer Service at 1-800-345-6443.
•
Because conditions and practices in health care institutions vary, 
this manual can only describe general guidelines. It is the user’s 
responsibility to ensure the validity and effectiveness of the 
methods used.
•
Because some environments cause a quicker collection of lint and 
dust than others, inspect and clean the fan filters more often than 
every 250 hours if necessary. 
•
The “Hardware” function and EST in the Diagnostics Mode should 
only be run by qualified personnel.
•
A “restart” is an infrequent event.
•
The gas return port on the ventilator is a cylindrical port which 
requires mating to a specified expiratory filter to seal the 
expiratory limb.
•
The humidifier power connection is available only on 100-120VAC 
ventilators.
•
Record O
2
sensor manufacturing or warranty numbers and
installation date for future reference. Save manufacturers 
instruction about end of life replacement.
•
To ensure accurate O
2
monitoring, check O
2
sensors periodically
and replace as per manufacturer specification.
•
Sensor performance and expected operating life information is 
outlined in the sensor manufacturer’s instructions for use. 
Thoroughly review all O
2
sensor instructions prior to installation
and use with the Respironics V200 Ventilator.
•
O
2
sensor calibration is performed during EST. If recalibration of
the O
2
sensor is required, follow the instructions in “Extended
Self Test (EST)” on page 11-4 for running EST.