REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

Appendix C-1

Appendix C. Alarm Testing Procedure

The following procedure is available if the operator wants to test the operation
of the following alarms. We recommend following “Preoperational Procedure”
on page 8-32 b
efore performing the Alarm Testing procedure. It is assumed
that the preoperational procedure has been run before the Alarm Testing
Procedure is followed.

Setup

1. Connect O

2

supply to the Respironics V200 Ventilator.

2. Connect optional O

2

sensor.

3. Run EST.

4. Attach the test lung to the patient wye. (Use the test lung provided

with your system.)

5. Use the following settings:

Mode VCV-A/C

Tidal Vol: 400ml

High Pressure (HIP): 50cmH

2

0 (hPa)

(or higher if required by the test lung)

Rate: 10 Bpm

Low Insp Press: 3 cmH

2

0 (hPa)

Peak Flow 40 Lpm

PEEP: 4 cmH

2

0 (hPa)

Low PEEP: 2 cmH

2

0 (hPa)

Low Vt Mand: 0 mL

PSV: 0 cmH

2

0 (hPa)

Low Vt Spont: 0 Lpm

I-Trigger: 2 cmH

2

0 (hPa)

High Rate: 150 Bpm

O

2

: 21%

Low VE: 1 L

Insp. Hold: 0 sec.

Apnea: 15 sec.

Apnea Rate: 20 Bpm