Apnea Alarm Test

1. Set the rate to 1 breath per minute.

2. Wait for 20 seconds.

3. The Respironics V200 should begin ventilating at a rate of 20 BPM

while activating the Apnea alarm.

4. Reset the rate to 10 BPM.

5. RESET the alarm.

6. The Respironics V200 should begin ventilating normally.