beautypg.com

Rs-232 configuration – Philips V200 User Manual

Page 203

background image

Options and Accessories

REF 1057983 A

Respironics V200 Ventilator Operator’s Manual

13-39

Chapter 13

RS-232 Communications Option 2 (Com2)

The Respironics V200 Ventilator allows the transmission of data from the
ventilator to the RS-232 communications interface. The Com2 option allows
the ventilator to communicate with data systems The ventilator has two
communications options, Com1 and Com2, which work exclusively of one
another.

RS-232 Configuration

When the Com2 option is installed, the RS-232 communications port is
configured for all communications functions in the following manner:

Baud Rate 9,600

Data Bits 8

Parity None

Stop Bits 1

NOTE:

In diagnostics mode, with Com2 installed, the baud rate is 19,200.

To properly use the Com2 option, connect the communications cable to the
9-pin serial connection on the back of the ventilator.

Figure 13-26: 9-pin serial port connection

WARNING:

It is the responsibility of the end user to validate the compatibility and use
of information transmitted from the ventilator with the device to be
connected to the ventilator.

WARNING:

The data provided via the Com2 option is for reference only. Decisions for
patient care should be based on the clinician’s observations of the patient.

Serial Port Connection