Calibrate touch probe via two points G748

Cycle G748 measures two opposite points and computes the touch

probe adjustment dimension and the ball diameter. If the tolerance

values defined in the cycle are exceeded, the cycle corrects the touch

probe data. The result of the measurement is saved additionally in the

variable #i99 (See "Touch probe cycles for automatic operation" on

page 431.).

Cycle run
From the current position the touch probe moves along the defined

measuring axis toward the measuring point. When the stylus touches

the workpiece, the measured value is saved and the touch probe is

positioned back to the starting point. For the pre-positioning for the

second measurement, the cycle first moves the touch probe by the

offset in the circumnavigation direction RB and then by the offset in the

measuring direction RC. The cycle performs the second probing

operation in the opposite direction and saves the result.
The control outputs an error message if the touch probe does not

reach any touch point within the defined measuring path. If a

maximum deviation WE was programmed, the measuring points are

approached twice and the mean value is saved as result. If the

difference of the measurements is greater than the maximum

deviation WE, the program run is interrupted and an error message is

displayed.

Example: G748 Calibrate touch probe via two
points

. . .

MACHINING

N3 G748 K20 AC10 EC33 Q0 P0 H0

. . .

Parameters

K

Incremental measuring path with direction (signed): Maximum

measuring path for probing. The algebraic sign determines the

probing direction.

RB Circumnavigation direction offset: Distance
RC Measuring direction offset: Distance for pre-positioning before

the second measurement

AC Nominal value for target position: Absolute touch point

coordinate

EC Nominal width: Coordinate for the second probing position
BE Tolerance width +/-: Range for the second measurement result

in which no compensation is applied

WE Maximum deviation: Probe twice and monitor the dispersion of

the measured values

F

Measuring feed rate: Feed rate for probing. If nothing is entered,

the measuring feed rate from the touch probe table is used. If

the entered measuring feed rate F is higher than the one in the

touch probe table, the feed rate is reduced to the value from the

touch probe table.

Q

Tool orientation: Orient the touch probe in the programmed

probing direction before each probing operation (machine-

dependent function)

P

PRINT outputs

0: OFF: Do not show measuring results

1: ON: Display measuring results