beautypg.com

Single-point measurement for zero point g771 – HEIDENHAIN SW 548328-05 DIN Programming User Manual

Page 435

background image

HEIDENHAIN MANUALplus 620, CNC PILOT 620/640

435

5.2 T

o

uc

h pr

obe cy

cles f

or single-point measur

ement

Single-point measurement for zero point G771

Cycle G771 measures with the programmed measuring axis in the

specified direction. If the tolerance value defined in the cycle is

exceeded, the cycle saves the measured deviation as zero point shift.

The result of the measurement is saved additionally in the variable

#i99 (See "Touch probe cycles for automatic operation" on page 431.).

Cycle run
From the current position the touch probe moves along the defined

measuring axis toward the measuring point. When the stylus touches

the workpiece, the measured value is saved and the touch probe is

positioned back to the starting point.
The control outputs an error message if the touch probe does not

reach any touch point within the defined measuring path. If a

maximum deviation WE was programmed, the measuring point is

approached twice and the mean value is saved as result. If the

difference of the measurements is greater than the maximum

deviation WE, the program run is interrupted and an error message is

displayed.

Example: G771 Single-point measurement for tool
compensation

. . .

MACHINING

N3 G771 R1 D0 K20 AC0 BD0.2 Q0 P0 H0

. . .

Parameters

R

Type of zero point shift:

1: Table and G59: Activate zero point shift and additionally

save in zero point table. The zero-point shift also remains

active after the program run.

2: Activate zero point shift with G59 for the further program

run. Zero point shift no longer active after program run.

D

Measuring axis: Axis in which the measurement is to be made

K

Incremental measuring path with direction (signed): Maximum

measuring path for probing. The algebraic sign determines the

probing direction.

AC Nominal value for target position: Touch point coordinate
BD Tolerance +/-: Measurement result range in which no

compensation is applied

WE Maximum deviation: Probe twice and monitor the dispersion of

the measured values

F

Measuring feed rate: Feed rate for probing. If nothing is entered,

the measuring feed rate from the touch probe table is used. If

the entered measuring feed rate F is higher than the one in the

touch probe table, the feed rate is reduced to the value from the

touch probe table.