DEAD TIME . . . . . . . . . . 0.1 to 100.0 sec

This value is defined as the time difference between a change in % power and the
start of an actual change in rate. The default value is 1.

Figure 4-8 Dead Time

MAXIMUM POWER . . . . 0.0 to 100.0%

This parameter is used to set the maximum permissible % power level. The control
voltage output will not exceed this limit. The default value is 100.0%.

DENSITY . . . . . . . . . . . . 0.50 to 99.99 g/cm

3

This parameter is specific to the material being deposited onto the crystal. It is one
of two parameters that relate the mass loading on the crystal to a thickness.
The default value is 1.00 g/cm

3

.

Z-Ratio . . . . . . . . . . . . . . 0.100 to 9.999

This parameter is specific to the material being deposited. It is one of two
parameters that relate the mass loading on the crystal to a thickness.
The default value is 1.000. See

Chapter A

for material density and Z-Ratio values.