INFICON XTC/3 Thin Film Deposition Controller Operating Manual User Manual
Page 66

3 - 20
PN
07
4-
44
6-
P1
J
XTC/3 Operating Manual
Ready (M only)
When each State starts, the old 
and new state output are set or 
cleared as necessary
When each State starts, the old and 
new state output are set or cleared 
as necessary
Crucib Switch 
(M only
Rise 1 (M only)
Soak 1 (M only)
Rise 2 (M only)
Soak 2
Shutter Delay 
(M only)
Deposit (M only)
Rate Ramp (M only)
Manual (M only)
Time Power (M only)
Idle Ramp (M only)
Idle (M only)
Max Power (M only)
When power reaches max power
When power is not at max power
Rate Dev Fail
(M only)
When the rate deviation is greater 
than 5% or 0.5Å of the desired 
rate for 60 seconds or 20 times 
the time constant if the time 
constant is greater than 3 
seconds
When rate deviation is within test 
limits.
When Idle Ramp is entered
Xtal Sw Fail
(M only)
When an attempted crystal switch 
fails
When another crystal switch is 
started
Xtal Switching
(M only)
Whenever a crystal switch is 
active
When the crystal switch is done
Crucib Sw Fail
(M only)
No crucible valid input within 30 
seconds of a crucible switch
When Start or Reset is done. 
If Stop on Alarms is set to No, XTC/3 
will give the error message but will 
not do anything else, it stays in 
Crucible Switch. If the Crucible Valid 
input becomes set, the output is 
cleared
Table 3-2 Output states (continued)
Output
Active
Inactive