2 standard query film name (xtc3/m only), 3 standard query film parameters, 4 standard query general parameter – INFICON XTC/3 Thin Film Deposition Controller Operating Manual User Manual
Page 125

5 - 13
PN
07
4-
44
6-
P1
J
XTC/3 Operating Manual
5.2.2.3.2 Standard Query Film Name (XTC3/M only)
Command:
QN
Film Number =
Response:
String = Null terminated series of ASCII characters
5.2.2.3.3 Standard Query Film Parameters
Command:
QF
Command ID =
Film Number =
Response:
Integer = 4 byte, low to high
Float = 4 byte, ANSII standard, single precision, low to high
Description: Type of response depends on command ID, see “DataType /
Format” column of
.
Special Case: QF255
given film, in numeric order.
5.2.2.3.4 Standard Query General Parameter
Command:
QG
Command ID =
Response:
Integer = 4 byte, low to high
Float = 4 byte, ANSII standard, single precision, low to high
Description: Type of response depends on command ID, see “DataType /
Format” column of
.
Special Case: QG255 will return all general parameters, in numeric order