2.3 Connecting the Controller

The operation of the XTC/3 depends on the proper connection of power and signal
interfaces to owner equipment and sources.

2.3.1 Verifying the Correct Input Voltage

WARNING - Risk Of Electric Shock

XTC/3 has line voltage present on the primary circuits
whenever it is plugged into a main power source.

Never remove the covers from XTC/3 during
normal operation.

There are no operator serviceable items within XTC/3.

Removal of the top or bottom covers must be done only
by a technically qualified person.

XTC/3 is powered by AC line current. The acceptable voltage input ranges and the
fuse rating and type are shown in

section 1.4.9 on page 1-10

.

2.3.2 Routing Oscillator (XIU) Cables

The signals traveling on this cable are both analog and digital. Do not route this
cable near areas with high levels of electromagnetic interference, even if its length
must be somewhat increased.

2.3.3 Interface Cable Fabrication and Pin-Out

It is necessary to fabricate several cables in order to interface the controller to the
deposition system. Refer to

section 2.2.3, Minimizing Noise Pickup From External

Cabling, on page 2-6

.

2.3.3.1 Source Control Connection

Two BNC connectors, labeled Source 1 and Source 2, provide the analog control
voltage to your source power supply. Standard 50 ohm coaxial cables may be used.