beautypg.com

Measurement Computing TempScan/1100 User Manual

Page 98

background image

5-8 System Operation

889897

TempScan / MultiScan User's Manual

Part 4 – The Trigger Block is now complete. The following describes a

U6

response when the Trigger event has

occurred, the Stop event has also occurred, and the End scan has also occurred. The Current Position of Read Pointer
field has the same value of the first available Pre-Trigger scan -100, and an additional 100 Post-Trigger scans plus 150
Post-Stop scans are available. The Trigger Block is now complete. The Time/Data stamping is in absolute ASCII
format.

0000001,0000351,-0000100,12:01:43.100,08/29/96,0000100,12:25:01.300,

08/29/96,0000250,01

Field

Field Name

Buffer Status String Response

Response Description

1

Number of Trigger Blocks Available

0000001

One

2

Number of Scans Available

0000351

100 + 1 + 100 + 150

3

Current Position of Read Pointer

-0000100

Pre-Trigger Scan -100

4

Time/Date Stamping of Trigger Event

12:01:43.100,08/29/96

Absolute ASCII format

5

Position of Stop Event Pointer

0000100

Post-Trigger Scan 100

6

Time/Date Stamping of Stop Event

12:25:01.300,08/29/96

Absolute ASCII format

7

Position of End Scan Pointer

0000250

Post-Trigger Scan 250

8

Status of Current Trigger Block

01

Acquisition complete


Query (U6) Buffer Operation – Single Trigger Block, Parts 3 & 4 (of 4)

Buffer

Buffer

Empty

Empty

Read Pointer >

Scan -100

Trigger Block
1 Pre-Trigger

Scans

Not

accessible Scan -100

Trigger Block
1 Pre-Trigger

Scans

Scan -099

Read Pointer >

Scan -099

Trigger Scan >

Scan 000

Trigger Block

1 Post-Trigger

Scans

Trigger

Scan

> Scan 000

Trigger Block

1 Post-Trigger

Scans

Scan 001

Scan 001

Stop Scan >

Scan 1004

Stop

Scan

>

Scan 1004

Scan 1005

Trigger Block

1 Post-Stop

Scans

Scan 1005

Trigger Block

1 Post-Stop

Scans

End Scan >

Scan 1254

End

Scan

>

Scan 1254

Block 2

Trigger Blocks

2 through 6

Block 2

Trigger Blocks

2 through 6

Block 3

Block 3

Block 4

Block 4

Block 5

Block 5

Scan 20216 >

Block 6

Scan 20215 >

Block 6

(1) The current Acquisition Buffer status.

(2) The oldest scan in the Acquisition

Buffer has been read.

This manual is related to the following products: