3-10 Scanning Cards

889897

TempScan / MultiScan User's Manual

MHV/24 High-Voltage Scanning Card

MHV/24 Specifications

Note: These specifications are subject to change without notice.

Number of Inputs: 24 differential; programmable by channel for any input range
Input Connectors: Screw Terminal.
Voltage Range, Accuracy and Resolution: Range is based on maximum peak-to-peak signal for AC volts.

Accuracy is based on AC voltages where the frequency of the input signal is an integer multiple of the
AC line cycle

± 1.0%, and with line-cycle integration enabled.

Range Accuracy

Resolution

± 2.5 V;

± 0.02% of range;

78.14

µV/bit

± 25 V;

± 0.02% of range;

781.4

µV/bit

± 250 V;

± 0.02% of range;

7.81 mV/bit

Digital Filtering: Averages 32 samples @ 50/60 Hz for line cycle noise rejection (DCV measurements).
Temperature Coefficient:

< 0.01%/°C.

Input Impedance: 10 MOhm typical.
Input Bias Current: 20 pA maximum.
Maximum Common Mode Voltage: 500 VDC or AC peak. Specified for coupling impedance > 30

MOhm, and common mode frequency < 60 Hz. Maximum 700 VDC or AC peak before equipment
damage occurs. Maximum 325 VDC or AC peak if used in the same system with the MTC/24 scanning
module.

Maximum Normal Mode Voltage: 500 VDC or AC peak.
Common Mode Rejection: 100 dB typical.
Maximum Allowable Input: 400 VDC.
Channel to Power Ground Isolation: 500 V peak.
Channel to Channel Isolation: 500 V peak. Maximum 325 VDC or AC peak if used in the same system

with the MTC/24 scanning module.

MHV/24 Description

The MHV/24 high-voltage scanning card contains 24 differential input channels and can measure analog
input signals on any one of the following three programmable voltage ranges: ±2.5V, ±25V, and ±250V.