Command summary, Ggain, Kkey – Measurement Computing TempScan/1100 User Manual
Page 202: Pchan, Smode, Tstate, Wmode, Achans, output, A#state, Cchans, type

A-6 API Command Reference
889897
TempScan / MultiScan User’s Manual
Command Summary
Command
Syntax
Description
@
Trigger On Command
@
Trigger on receipt of an
@
.
*B
Flush Acquisition
Buffer
*B
Flush any data currently in the Acquisition Buffer.
*C
Clear Channel
Configuration
*C
Clear out the channel configuration so that no channels are
configured.
*F
Restore Factory
Settings
*F
Restore unit to factory configuration.
*Ggain
Set RTD Gain Calibration reference point to specified number
of ohms. gain is the reference point in the form of a real
number nnn.nn
*G
Set RTD Gain
Calibration
Reference
(TempScan/1100 only)
G?
Query the present reference point setting
*K
Change Calibration
Keyword
*Kkey
Change the calibration keyword where key is the keyword in
the form of a 5-digit number nnnnn. Therefore the
password can be no longer than 5 characters.
*P
Adjust Calibration Card
Pots
*Pchan
Select voltage level to be enabled for adjustment. Valid
options for chan are 01 – Enable –4.4 Volt Source on
calibration card. 02 – Enable +4.4 Volt Source on calibration
card. and 03 – Select 0 Volts Ground.
*R
Reset Power-On
*R
Power-on reset. Equivalent to a hardware reset.
*S
Set Power-Up
Configuration
*Smode
Specify power-up configuration mode. Valid options for mode
are 0 – Power-up under last known configuration (default)
and 1 – Power-up under factory default configuration.
*T
Set Scan Time
Stamping
*Tstate
Specify whether or not scan time stamping will be enabled.
Valid options for state are: 0 – Disable scan time stamping
(default), 1 – Enable absolute time stamping, and 2 – Enable
relative time stamping.
*W
Set Software Digital
Filtering
(TempScan/1100
only)
*Wmode
Specify whether or not software digital filtering for
thermocouple cards will be enabled. Valid options for mode
are: 0 – Disable digital filtering for thermocouple cards, and
1 – Enable digital filtering for thermocouple cards (default)
Achans,
output
Set alarm condition where chans is the channel(s) to assign to
the digital output.
chans may be in two forms: chan (for a single channel where
1
[[[[chan [[[[ max) or first – last (for a range of channels where
1
[[[[ first < last [[[[ max), where max = 992 for the
TempScan/1100 and max = 744 for the MultiScan/1200.
output is the digital output number to associate the channel(s)
with, where 1<=output<=32, and 0 is null.
A
Assign Alarm Output
A?
Query channels assigned to digital output.
A#
Set Scan Alarm
Stamping
A#state
Specify whether or not scan alarm stamping will be enabled.
Valid options for state are 0 – Disable scan alarm stamping
(default), and 1 – Enable scan alarm stamping.
Cchans,
type
[,
[lowsp],[
highsp],
[hyst] ]
Configure the channels for the master unit and its slave units.
chans and type are required arguments, while lowsp,
highsp and hyst are optional.
chans may be in two forms: chan (for a single channel where 1
[[[[ chan [[[[ max) or first – last (for a range of channels where
1
[[[[ first < last [[[[ max), where max = 992 for the
TempScan/1100 and max – 744 for the MultiScan/1200.
The chassis is chan = 0.
type is the type to assign to those channels. 0 is null.
C
Configure Channels
C?
Query current channel configuration