Calibration of master chassis – Measurement Computing TempScan/1100 User Manual

Page 120

background image

6-4 System Calibration

879596

TempScan / MultiScan User's Manual

Calibration of Master Chassis

Note: The following text discusses the manual method of calibration. However, we strongly

suggest using ChartView’s internal automated calibration method. To do so, from
ChartView navigate as follows, then follow the on screen instructions.

Device Select Status Calibration [then follow on screen prompts]

Note: Prior to each calibration session, disable calibration protection by setting the rear panel

DIP microswitch 9 to the up (1) position. Then enable Calibration Mode via the Enter
Calibration Mode
(

K

) command with a valid password. If the default calibration

password has not been changed, enter

K12345

.

Note: At the end of each calibration session, disable Calibration Mode with the End Calibration

Mode (

E

) command. Then re-enable calibration protection by resetting DIP

microswitch 9 to the down (0) position.