Calibration of master chassis – Measurement Computing TempScan/1100 User Manual
Page 120

6-4 System Calibration
879596
TempScan / MultiScan User's Manual
Calibration of Master Chassis
Note: The following text discusses the manual method of calibration. However, we strongly
suggest using ChartView’s internal automated calibration method. To do so, from
ChartView navigate as follows, then follow the on screen instructions.
⇒ Device ⇒ Select Status ⇒ Calibration [then follow on screen prompts]
Note: Prior to each calibration session, disable calibration protection by setting the rear panel
DIP microswitch 9 to the up (1) position. Then enable Calibration Mode via the Enter
Calibration Mode (
K
) command with a valid password. If the default calibration
password has not been changed, enter
K12345
.
Note: At the end of each calibration session, disable Calibration Mode with the End Calibration
Mode (
E
) command. Then re-enable calibration protection by resetting DIP
microswitch 9 to the down (0) position.