Software Reference

988992

ChartView , CV-29

Device

… Status ¾

Instrument

Inventory

Accesses the System Inventory dialog box.

System Inventory Dialog Box

Alarms

Brings up selection list for Status, Log configuration, and Log Enable

Status

Brings up the Alarm Status dialog box, which shows the
current state of each alarm. The alarm state appears as
either a “1” or “0”, the “1” being digital High, and the “0”
being digital Low (0). A one-for-one correlation between
the alarm number and channel can only occur if each
channel of a 32 channel [or less] system is assigned to a
dedicated alarm number.

Alarm Status Dialog Box

Log configuration

Alarm logging is a feature of ChartView Plus only. Alarm Logging allows you to monitor the
Alarm states of the instrument, and save selected information to a file whenever any of the states
change.

Alarm logging is implemented by periodically polling the instrument; the
time interval between polls is about one second. With this in mind, be
aware any event that could trigger an alarm [lasting one second or less
],
could be missed and not logged.

Alarm Log Configuration Dialog Box