2-8 Expansion Units

889897

TempScan / MultiScan User's Manual

Two (2) switches, three (3) connectors, one (1) grounding nut, and ten (10) input card slots on the rear
panel of the Exp/11A provide power, slave addressing, master/slave connections, a single point grounding
node, and scanning card expansion.

Power Switch: Used to turn power to the Exp/11A ON and OFF. When the switch is in the depressed

position the power is ON. When in the extended position, the power is OFF.

DIP Switch: Used for selecting the Exp/11A slave address ID.
Power Connector: Provides power for the unit. Internally configurable for either 105-125 or 210-250

VAC, 50/60Hz, plus fuse circuit breaker.

Master/Slave Connector: Two DB25 master/slave ports provide one connection to a TempScan/1100,

MultiScan/1200, Exp/10A or Exp/11A unit, and one connection to another Exp/10A or Exp/11A
expansion unit.

Grounding Screw: An external single-point grounding node has been supplied for (but not limited to)

thermocouple shield termination.

Shielded Enclosure: For the TempScan/1100: Accepts any combination of TempTC/32B,

TempV/32B, and/or TempRTD/16B scanning cards. For the MultiScan/1200: Accepts any
combination of MTC/24 and/or MHV/24 scanning cards. Note that the TempScan/1100 and
MultiScan/1200 scanning cards must not be mixed within the same system.

Exp/11A Specifications

CAUTION

CAUTION

CAUTION

CAUTION

Please read this manual carefully! If equipment is used in any manner not
specified in this manual, the protection provided by the equipment may be
impaired.

Note: These specifications are subject to change without notice.

Installation Category: For CE: Category 2 for Line Voltage Input terminal. All other terminals are

Category 1.

Master/Slave Port: Female DB25 (2).
Number of Slots: Ten (10).
Number of Channels (TempScan/1100): Up to 320 differential voltage or thermocouple inputs, or up to

160 RTD inputs; accepts any combination of ten (10) TempTC/32B, TempV/32B or TempRTD/16B
scanning modules.

Number of Channels (MultiScan/1200): Up to 240 differential voltage or thermocouple inputs; accepts

any combination of ten (10) MTC/24 and MHV/24 scanning modules.

Channel to System Isolation: 60V peak.
Dimensions: 425mm wide x 305mm deep × 135mm high (16.75” × 12” × 5.75”).
Weight: 6.36 kg. (14 lbs.).
Operating Environment: For standard: Indoor use, 0 to 50°C; 0 to 95% RH (non-condensing) to 35°C;

linearly derate 3% RH/°C from 35 to 50°C; For CE: Indoor use at altitudes below 2000 m,
0 to 40

°°°°C; 0 to 80% RH up to 31°°°°C decreasing linearly 4% RH/°°°°C to 40°°°°C.

Controls: Power Switch (external), DIP switch for setting slave ID (external).
Front Panel Indicators:
LED indicators for SCAN, ERROR, and POWER.
Power: 105-125 or 210-250 VAC, 50/60 Hz; 20 VA maximum

CAUTION

CAUTION

CAUTION

CAUTION

Line Voltage: The protective conductor terminal on the AC line connector must
be connected to an external protective earthing system. Failure to make such
a connection will impair protection from shock.