Caution – Measurement Computing TempScan/1100 User Manual
Page 125

TempScan / MultiScan User's Manual
879596
System Calibration 6-9
To Calibrate a Thermocouple Card for Gain
CAUTION
Ensure that the thermocouple scanning card has been calibrated for offset. If not,
follow the thermocouple card offset calibration procedure as described in the
previous text.
1. Connect a precision voltage source to Channel 1 on the scanning card being calibrated. Use standard
copper hookup wire. Do not use thermocouple wire. Slide the scanning card into the previously-
calibrated master chassis.
2. Power up the unit and wait at least one hour before continuing.
3. Set the Part A precision voltage source and send its corresponding Calibrate Channel Gain (
G
)
command
Gchan,type
– where
chan
refers to the first channel number on the scanning card being
calibrated (refer to the previous table for valid gain channel numbers) – from the following list:
• Part A: -13.82 mV
Gchan,7
• Part I:
-47.70 mV
Gchan,8
• Part B: +13.82
mV
Gchan,7
• Part J:
+47.70 mV
Gchan,8
• Part C: -18.70 mV
Gchan,6
• Part K: -54.90 mV
Gchan,2
• Part D: +18.70
mV
Gchan,6
• Part L: +54.90 mV
Gchan,2
• Part E: -21.12 mV
Gchan,3
• Part M: -77.00 mV
Gchan,4
• Part F: +21.12
mV
Gchan,3
• Part N: +77.00 mV
Gchan,4
• Part G: -43.00 mV
Gchan,1
• Part O: -100.00 mV
Gchan,10
• Part H: +43.00
mV
Gchan,1
• Part P: +100.00 mV
Gchan,10
Wait for the Part A command to complete. The TRIGGER LED indicator will flash for a few seconds.
4. Monitor the Calibration Status Register (
CSR
) to ensure that the Calibrate Channel Gain (
G
) command
completed without error. Refer to the User Status (
U
) command
U2
to monitor and decode the
Calibration Status Register (
CSR
).
5. Repeat Steps 3 and 4 for each of the next fifteen parts: Part B through Part P.