beautypg.com

Measurement Computing TempScan/1100 User Manual

Page 96

background image

5-6 System Operation

889897

TempScan / MultiScan User's Manual

Note: The responses to all queries (including

U6

,

R1

, and

R2

) can be configured with specific field

separators and response terminators. For details on setting query terminators, see command
Set Query Terminator (

Q

) in the chapter API Command Reference.

Buffer

Buffer

N/A Trigger

Block

1 Pre-Trigger

Scans

Read Pointer >

Scan -100

Trigger Block
1 Pre-Trigger

Scans

N/A

Scan -099

N/A

Scan -098

Trigger Scan >

N/A

Trigger Block

1 Post-Trigger

Scans

Trigger

Scan

> Scan 000

Trigger Block

1 Post-Trigger

Scans

N/A

Scan 001

N/A

Write Pointer >

Scan 050

Stop Scan >

N/A

Stop Scan >

N/A

N/A Trigger

Block

1 Post-Stop

Scans

N/A

Trigger

Block

1 Post-Stop

Scans

N/A

N/A

End Scan >

N/A

End Scan >

N/A

(1) The Acquisition Buffer is empty.

(2) The Stop event has not yet occurred.

Part 1 – The Acquisition Buffer is empty. The following describes a Query the Buffer Status String (

U6

) response

when the Acquisition Buffer is empty. The Time/Data stamping is in absolute ASCII format.

0000000,0000000,-0999999,00:00:00.000,00/00/00,-0999999,00:00:00.000, 00/00/00,-

0999999,00

Field

Field Name

Buffer Status String Response

Response Description

1

Number of Trigger Blocks Available

0000000

Zero

2

Number of Scans Available

0000000

Zero

3

Current Position of Read Pointer

-0999999

Undefined

4

Time/Date Stamping of Trigger Event

00:00:00.000,00/00/00

Absolute ASCII format

5

Position of Stop Event Pointer

-0999999

Undefined

6

Time/Date Stamping of Stop Event

00:00:00.000,00/00/00

Absolute ASCII format

7

Position of End Scan Pointer

-0999999

Undefined

8

Status of Current Trigger Block

00

Not being acquired

This manual is related to the following products: