TempScan / MultiScan User's Manual

879596

System Calibration 6-5

To Calibrate the Master Chassis for Offset

1. For offset calibration, locate the calibration jumpers located in one corner of an appropriate scanning

card, remove the two calibration jumper caps and short the calibration jumpers using one of those
jumper caps, as shown in the figure. Slide the scanning card into the master chassis.

2. Power up the unit and wait at least one hour before continuing.

3. Send the Calibrate Channel Offset (

H

) command

H0

– where channel

0

refers to the chassis – to the

unit, and wait for the command to complete. The TRIGGER LED indicator will flash for a few
seconds.

4. Monitor the Calibration Status Register (

CSR

) to ensure that the Calibrate Channel Offset (

H

)

command completed without error. Refer to the User Status (

U

) command

U2

to monitor and decode

the Calibration Status Register (

CSR

).

5. To monitor the chassis calibration voltage, configure a normal acquisition with Channel 3 set as

channel type 90. The input will be returned in Counts.

To Calibrate the Master Chassis for Gain

CAUTION

Ensure that the master chassis has first been calibrated for offset. If not, follow
the master chassis offset calibration procedure as described in the previous text.

1. For gain calibration, slide the scanning card out from the chassis, locate the calibration jumpers,

remove any calibration jumper caps, and connect clip leads from the precision voltage calibration
source, as shown in the figure. Use standard copper hookup wire. Do not use thermocouple wire.
Slide the scanning card into the master chassis.

2. Power up the unit and wait at least one hour before continuing.

3. Set the precision voltage source to -4.4 (negative) Volts DC.

4. Send the Calibrate Channel Gain (

G

) command

G0,91

to the unit – where channel

0

refers to the

chassis, and channel type

91

refers to a reserved calibration type – and wait for the command to

complete. The TRIGGER LED indicator will flash for a few seconds.

5. Monitor the Calibration Status Register (

CSR

) to ensure that the Calibrate Channel Gain (

G

) command

completed without error. Refer to the User Status (

U

) command

U2

to monitor and decode the

Calibration Status Register (

CSR

).

6. To monitor the chassis calibration voltage, configure a normal acquisition with Channel 1 set as

channel type 90. The input will be returned in Counts.

7. Next, set the precision voltage source to +4.4 (positive) Volts DC.

8. Repeat Steps 4, 5, and 6.