TempScan / MultiScan User's Manual

889897

System Operation 5-7

Part 2 – The Stop event has not yet occurred. The following describes a

U6

response when the Trigger event has

occurred but the Stop event has not yet occurred. The Current Position of Read Pointer field has the value of the first
available Pre-Trigger scan -100, and an additional 50 Post-Trigger scans are available. The Time/Data stamping is in
absolute ASCII format.

0000001,0000151,-0000100,12:01:43.100,08/29/96,-0999999,00:00:00.000, 00/00/00,-

0999999,00

Field

Field Name

Buffer Status String Response

Response Description

1

Number of Trigger Blocks Available

0000001

One

2

Number of Scans Available

0000151

100 + 1 + 50

3

Current Position of Read Pointer

-0000100

Pre-Trigger Scan -100

4

Time/Date Stamping of Trigger Event

12:01:43.100,08/29/96

Absolute ASCII format

5

Position of Stop Event Pointer

-0999999

Undefined

6

Time/Date Stamping of Stop Event

00:00:00.000,00/00/00

Absolute ASCII format

7

Position of End Scan Pointer

-0999999

Undefined

8

Status of Current Trigger Block

00

Still being acquired


Query (U6) Buffer Operation – Single Trigger Block, Parts 1 & 2 (of 4)

Buffer

Buffer

Read Pointer >

Scan -100

Trigger Block
1 Pre-Trigger

Scans

Read Pointer >

Scan -100

Trigger Block
1 Pre-Trigger

Scans

Scan -099

Scan -099

Scan -098

Scan -098

Trigger Scan >

Scan 000

Trigger Block

1 Post-Trigger

Scans

Trigger

Scan

> Scan 000

Trigger Block

1 Post-Trigger

Scans

Scan 001

Scan 001

Scan 050

Scan 050

Stop Scan >

Scan 100

Stop

Scan

>

Scan 100

Scan 101

Trigger Block

1 Post-Stop

Scans

Scan 101

Trigger Block

1 Post-Stop

Scans

Write Pointer >

Scan 150

Scan 150

End Scan >

N/A

End Scan >

Scan 250

(3) The End scan has not yet occurred.

(4) The Trigger Block is now complete.

Part 3 – The End scan has not yet occurred. The following describes a

U6

response when the Trigger event has

occurred, the Stop event has also occurred, but the End scan has not yet occurred. The Current Position of Read
Pointer field has the same value of the first available Pre-Trigger scan -100, but now an additional 100 Post-Trigger
scans plus 50 Post-Stop scans are available. The Time/Data stamping is in absolute ASCII format.

0000001,0000251,-0000100,12:01:43.100,08/29/96,0000100,12:25:01.300, 08/29/96,-

0999999,00

Field

Field Name

Buffer Status String Response

Response Description

1

Number of Trigger Blocks Available

0000001

One

2

Number of Scans Available

0000251

100 + 1 + 100 + 50

3

Current Position of Read Pointer

-0000100

Pre-Trigger Scan -100

4

Time/Date Stamping of Trigger Event

12:01:43.100,08/29/96

Absolute ASCII format

5

Position of Stop Event Pointer

0000100

Post-Trigger Scan 100

6

Time/Date Stamping of Stop Event

12:25:01.300,08/29/96

Absolute ASCII format

7

Position of End Scan Pointer

-0999999

Undefined

8

Status of Current Trigger Block

00