Calibrate channel offset – Measurement Computing TempScan/1100 User Manual

Page 232

background image

A-36 API Command Reference

889897

TempScan / MultiScan User’s Manual

H

- Calibrate Channel Offset

TYPE

Calibration

EXECUTION

Immediate

SYNTAX

Hchan

Calibrate the correction offset constants for given option card where

chan

is

the channel number such that

1

<

chan

<

max

, where

max

=

992

for the

TempScan/1100 and

max

=

744

for the MultiScan/1200. The chassis is

chan = 0

.

DESCRIPTION

Note: This command is only for advanced users who perform their own calibration. It is not necessary

for normal, everyday operation.

Active only in Calibration Mode, the Calibrate Channel Offset (

H

) command is used to calibrate the correction

offset constants for a given scanning card. The channel offset values are necessary in order for the internal
firmware to accurately interpret readings coming from the input channels. This command will take the parameter

chan

and calculate the offsets based on the present channel input voltage.

For more information on manual calibration, see chapter System Calibration.

Note: To achieve correct results, a short must be connected across the desired channel before performing this

command. After the command has been issued, the precision voltage source must remain at its value at
least until the TRIGGER LED indicator stops flashing.


EXAMPLE

PRINT#1, “OUTPUT07;K12345X”

‘ Enter Calibration Mode

‘ Connect short across the desired channel

PRINT#1, “OUTPUT07;H524X”

‘ Calculate offsets for slave card 16 (TempScan/1100)

‘ Wait for command to complete

‘ Remove precision voltage

PRINT#1, “OUTPUT07;EX”

‘ End Calibration Mode