System Calibration 6-11

4. Send the following Calibrate Cold Junction Offset (

J

) command

Jchan,type,temp

for the first cold-

junction channel (temperature sensor #1):

• TempTC/32B (for TempScan/1100):

Jchan,type,nnn.n

• TempTC/32A (for TempScan/1000A):

Jchan,type,nnn.n

• MTC/24 (for MultiScan/1200):

Jchan,type,nnn.n

Wait for the command to complete. The TRIGGER LED indicator will flash for a few seconds.

Monitor the Calibration Status Register (

CSR

) to ensure that the Calibrate Cold Junction Offset (

J

)

command completed without error. Refer to the User Status (

U

) command

U2

to monitor and decode

the Calibration Status Register (

CSR

).

5. Next, send the following Calibrate Cold Junction Offset (

J

) command

Jchan,type,temp

for the

second cold-junction channel (temperature sensor #2):

• TempTC/32B (for TempScan/1100):

Jchan+4,type,ppp.p

• TempTC/32A (for TempScan/1000A):

Jchan+4,type,ppp.p

• MTC/24 (for MultiScan/1200):

Jchan+8,type,ppp.p

Wait for the command to complete. The TRIGGER LED indicator will flash for a few seconds.

Monitor the Calibration Status Register (

CSR

) to ensure that the Calibrate Cold Junction Offset (

J

)

command completed without error. Refer to the User Status (

U

) command

U2

to monitor and decode

the Calibration Status Register (

CSR

).

6. Next, send the following Calibrate Cold Junction Offset (

J

) command

Jchan,type,temp

for the

third cold-junction channel (temperature sensor #3):

• TempTC/32B (for TempScan/1100):

Jchan+16,type,qqq.q

• TempTC/32A (for TempScan/1000A):

Jchan+16,type,qqq.q

• MTC/24 (for MultiScan/1200):

Jchan+16,type,qqq.q

Wait for the command to complete. The TRIGGER LED indicator will flash for a few seconds.

Monitor the Calibration Status Register (

CSR

) to ensure that the Calibrate Cold Junction Offset (

J

)

command completed without error. Refer to the User Status (

U

) command

U2

to monitor and decode

the Calibration Status Register (

CSR

).

7. At this point, the 3 cold-junction channels of the MTC/24 (for MultiScan/1200) scanning card have

been calibrated. This additional Step applies only to the fourth cold-junction channel of the
TempTC/32B (for TempScan/1100) and TempTC/32A (for TempScan/1000A) scanning cards.

Finally, send the following Calibrate Cold Junction Offset (

J

) command

Jchan,type,temp

for the

fourth cold-junction channel (temperature sensor #4):

• TempTC/32B (for TempScan/1100):

Jchan+20,type,rrr.r

• TempTC/32A (for TempScan/1000A):

Jchan+20,type,rrr.r

Wait for the command to complete. The TRIGGER LED indicator will flash for a few seconds.

Monitor the Calibration Status Register (

CSR

) to ensure that the Calibrate Cold Junction Offset (

J

)

command completed without error. Refer to the User Status (

U

) command

U2

to monitor and decode

the Calibration Status Register (

CSR

).