6-10 System Calibration

879596

TempScan / MultiScan User's Manual

To Calibrate a Thermocouple Card for Cold-Junction Offset

CAUTION

Ensure that the thermocouple scanning card has been calibrated for offset and
gain. If not, follow the thermocouple card offset and gain calibration procedures
as described in the previous text.

Cold-Junction Offset Calibration


1. Connect accurate thermocouple sources to the following multiple channels on the scanning card being

calibrated, where

chan

refers to the first channel number on the scanning card being calibrated (refer

to the previous table for valid cold-junction channel numbers).

• TempTC/32B (for TempScan/1100):

chan, chan+4, chan+16, chan+20

• TempTC/32A (for TempScan/1000A):

chan, chan+4, chan+16, chan+20

• MTC/24 (for MultiScan/1200):

chan, chan+8, chan+16

Slide the scanning card into the previously-calibrated master chassis.

2. Power up the unit and wait at least one hour before continuing.

3. Apply the following temperature values in Celsius to the multiple cold-junction channels above:

• TempTC/32B (for TempScan/1100):

nnn.n, ppp.p, qqq.q, rrr.r

• TempTC/32A (for TempScan/1000A):

nnn.n, ppp.p, qqq.q, rrr.r

• MTC/24 (for MultiScan/1200):

nnn.n, ppp.p, qqq.q