Memory allocation, Measuring modes (multiscan/1200 only) – Measurement Computing TempScan/1100 User Manual

Page 60

background image

4-2 System Configuration

879596

TempScan / MultiScan User's Manual

Memory Allocation

Each TempScan/1100 and MultiScan/1200 unit comes equipped with standard 256 KB memory. This is
expandable to 8 MB (TempMEM8 or MultiMEM8).

When expanded memory is added, the TempScan/1100 or MultiScan/1200 unit will automatically allocate
its memory for optimal use according to its current configuration. TempScan/1100 or MultiScan/1200
memory is broken into the following two main regions:
High/Low/Last (HLL) Registers (only configured channels)
Acquisition Buffer

When channels are configured, the TempScan/1100 or MultiScan/1200 calculates how much memory will
be needed for the High/Low/Last (HLL) Registers. The memory allocated to the HLL Registers would be
the following value –

(Number of channels) x 20 bytes

– while the remaining memory would be

allocated to the Acquisition Buffer. For example, if the memory option is 256K (

262144 bytes

) and the

number of configured channels is 96, then the amount allocated to the Acquisition Buffer would be:

262144 bytes Total Memory (256 KB)
- 1920 bytes High/Low/Last (96 x 20)
260224 bytes Acquisition Buffer Allocation (254.125 KB)

Measuring Modes (MultiScan/1200 Only)

The operation of the MultiScan/1200 unit involves two basic internal measuring modes, as follows:
Line-Cycle Integration / High-Speed Multi-Channel (or Default) Mode. In the line-cycle integration

/ high-speed multi-channel mode, the MultiScan/1200 samples channel data at 1.92 kHz or once every
520.83 µs. You can select the number of samples over which to average in order to accomplish noise
filtering.

Single-Channel High-Speed Burst Mode. In the single-channel burst mode, the MultiScan/1200

collects data in user-defined multiples of 256 samples at a user-defined sampling frequency between
38.5 Hz and 20.0 kHz.


The following commands are provided to setup the MultiScan/1200 for a given measuring mode.

M#mode

: Select the measuring mode. The factory default is line cycle integration / high-speed multi-

channel mode.

W#wt

: Select the number of samples over which to average for each channel. This is valid only in line

cycle integration / high-speed multi-channel mode. The factory default is 32 samples.

F#freq

: Select the burst mode frequency. This is valid only in single-channel high-speed burst

mode. The factory default is 20 kHz (or 20000 Hz).

Y0,post,0

: In line-cycle integration / high-speed multi-channel mode, select the number of Post-

Trigger scans.

In single-channel high-speed burst mode, select the number of 256-sample blocks to acquire. The
number must be a power of 2.

U16

: Query the measuring mode parameters

M#

,

W#

,

F#

, and/or

Y

to verify your selections.


For more information on the above commands, see Appendix A - API Command Reference.