Expanded Memory Configuration

The TempScan/1100 or MultiScan/1200 can accommodate 8 MBytes of memory, which is ideal for both
high-speed and long-term data logging. Measurements can be stored in memory and read out by
a controlling computer as time permits. Readings may be transferred at up to 300 KBytes per second over
the IEEE 488 bus, or up to 9600 baud using the unit’s standard RS-232 port.

To install expanded memory into the unit, it is necessary to perform the following steps:

WARNING

Never disassemble the case while it is connected to the AC power line!

Internal voltage potentials exist which could cause bodily injury or death!

WARNING

Never disconnect the AC power line from the TempScan/1100 or MultiScan/1200

while its scanning cards are connected to an external device! Common mode
voltage potentials exceeding 60 VDC or 30 Vrms at the terminals, may exist which
could cause bodily injury or death!

Note: If disassembly or disconnections are necessary, first turn off the power, then disconnect the

scanning cards, next disconnect the AC power line, and then any other cables, prior to
disassembly.

To Install Expanded Memory

1. Turn off the power, disconnect the scanning cards, the power line cord, and then all other cables

from the unit.

2. Place the unit on a flat surface. Remove the six screws on top of the case and remove the top cover.

3. Located on the main circuit-board assembly, alongside the scanning card enclosure, are SIMM

memory module slots JP8 and JP7. Remove the 256 KB module from JP7, insert one
4 MB module in its place, and insert the second 4 MB module in JP8.

4. Carefully reassemble the unit, replacing the top cover and screws.

Note: For re-assembly, first reconnect the AC power line (with the power OFF), next reconnect the

scanning cards, and then any other cables, prior to reapplying power to the entire system.