4-4 System Configuration

879596

TempScan / MultiScan User's Manual

Note: The MultiScan/1200 unit only supports the measurement of AC voltages where the frequency

of the input signal is an integer multiple of the AC line cycle. Consequently, sampling
weights less than 32 can yield incorrect results. In order to compute an accurate Vrms value,
you must program a frequency which will yield a sufficient even integer number of samples
per line cycle.

To Enable the Default Mode

To enable line-cycle integration / high-speed multi-channel mode, issue the following commands:

M#mode

: Select line-cycle integration / high-speed multi-channel mode, where

mode

is

0

.

W#wt

: Select the number of samples over which to average, where

wt

is the one of the following

weights:

1

,

2

,

4

,

8

,

16

,

32

(default),

64

,

128

, and

256

.

Y0,post,0

: Select the number

post

of Post-Trigger scans.

Cchans,type

: Select the channel number or range, and the scanning-card channel type(s).

Tstart,stop,0,0

: Set the Trigger (trigger start event) to the software Trigger On (

@

) command,

where

start

is

1

; and set the Stop (trigger stop event) to Count, where

stop

is

8

. The Set Counts (

Y

)

command is made valid when the Stop is set to Count.

@

: Start the acquisition.

For more information on the above commands, see Appendix A- API Command Reference.

Common Mode Voltage Inaccuracies

Inaccuracies can result if common mode voltages on adjacent channels are widely dissimilar. This
inaccuracy is due to inadequate settling time at the instrumentation amplifier when the unit is scanning
between channels. To eliminate these inaccuracies, the settling time can be increased using the Set Relay
Make Time
(

D#

) command. For more information on this command, see Appendix A-

API Command

Reference

.