TempScan / MultiScan User’s Manual

899493

System Overview 1-19

CAUTION

It is possible to place the line voltage switch (S2 or SW2) in a partial position

which could cause equipment damage or malfunction. When changing the
position of the line voltage selection switch (S2 or SW2), make sure the switch is
completely positioned to the 115 V or 220 V selection. The switch will “click” into
place when properly positioned.

4. Install a power line fuse appropriate for the line voltage. See section Fuse Replacement – Step 3,

following this section.

CAUTION

Do not use a fuse with a rating higher than specified. Otherwise the unit may be

damaged. If the instrument repeatedly blows fuses, locate and correct the cause of
the trouble before replacing the fuse.

5. Make note of the new voltage setting for later reference.
6. Carefully reassemble the unit.

Note: For re-assembly, first reconnect the AC power line (with the power OFF), next reconnect the

scanning cards, and then any other cables, prior to reapplying power to the entire system.