TempScan / MultiScan User’s Manual
889897
API
Command Reference A-17
*K
- Change Calibration Keyword
TYPE
Calibration
EXECUTION
Immediate
SYNTAX
*Kkey
Change the calibration keyword, where
key
is the keyword in the form of a 5-
digit number
nnnnn
. Therefore the password can be no longer than 5
characters.
DESCRIPTION
Note: This command is only for advanced users who perform their own calibration. It is not necessary
for normal, everyday operation.
The Change Calibration Keyword (
*K
) command allows the changing of the calibration keyword which is a user-
defined password. This password is a safety feature to prevent unauthorized personnel from entering Calibration
Mode and potentially altering calibration constants. The new keyword must be a 5-digit decimal number. Once
the keyword is set, it will thereafter be used by the Enter Calibration Mode (
K
) command to determine entry into
Calibration Mode.
For more information on manual calibration, see chapter System Calibration.
Note: This command is only valid when the calibration protection is disabled via the rear DIP microswitch 9.
EXAMPLE
PRINT#1, “OUTPUT07;K12345X”
‘ Enter Calibration Mode. Keyword =
12345
PRINT#1, “OUTPUT07;U2X”
‘ Check for keyword error
PRINT#1, “ENTER07”
INPUT#2, C%
‘ Get the Calibration Status Register (
CSR
)
IF (C% AND 16) THEN
PRINT,"ERROR — INVALID KEYWORD”
SLEEP 5
‘ If so, wait until keyword check jumper is disabled
PRINT#1, “OUTPUT07;*K67890X”
‘ Enter new Keyword =
67890
ENDIF
PRINT#1, “EX”
‘ End Calibration Mode