TempScan / MultiScan User’s Manual

889897

API

Command Reference A-17

*K

- Change Calibration Keyword

TYPE

Calibration

EXECUTION

Immediate

SYNTAX

*Kkey

Change the calibration keyword, where

key

is the keyword in the form of a 5-

digit number

nnnnn

. Therefore the password can be no longer than 5

characters.

DESCRIPTION

Note: This command is only for advanced users who perform their own calibration. It is not necessary

for normal, everyday operation.

The Change Calibration Keyword (

*K

) command allows the changing of the calibration keyword which is a user-

defined password. This password is a safety feature to prevent unauthorized personnel from entering Calibration
Mode
and potentially altering calibration constants. The new keyword must be a 5-digit decimal number. Once
the keyword is set, it will thereafter be used by the Enter Calibration Mode (

K

) command to determine entry into

Calibration Mode.

For more information on manual calibration, see chapter System Calibration.

Note: This command is only valid when the calibration protection is disabled via the rear DIP microswitch 9.

EXAMPLE

PRINT#1, “OUTPUT07;K12345X”

‘ Enter Calibration Mode. Keyword =

12345

PRINT#1, “OUTPUT07;U2X”

‘ Check for keyword error

PRINT#1, “ENTER07”

INPUT#2, C%

‘ Get the Calibration Status Register (

CSR

)

IF (C% AND 16) THEN

PRINT,"ERROR — INVALID KEYWORD”

SLEEP 5

‘ If so, wait until keyword check jumper is disabled

PRINT#1, “OUTPUT07;*K67890X”

‘ Enter new Keyword =

67890

ENDIF

PRINT#1, “EX”

‘ End Calibration Mode