beautypg.com

End calibration mode – Measurement Computing TempScan/1100 User Manual

Page 226

background image

A-30 API Command Reference

889897

TempScan / MultiScan User’s Manual

E

- End Calibration Mode

TYPE

Calibration

EXECUTION

Immediate

SYNTAX

E

Terminate Calibration Mode

DESCRIPTION

Note: This command is only for advanced users who perform their own calibration. It is not necessary

for normal, everyday operation.

To use this command the unit must be in Calibration Mode. The End Calibration Mode (

E

) command is then used

to take the unit out of Calibration Mode. Once this command has been executed the unit will no longer be in
Calibration Mode and the Enter Calibration Mode (

K

) command must be issued to re-enter Calibration Mode.

For more information on manual calibration, see chapter System Calibration.

EXAMPLE

PRINT#1, “OUTPUT07;K12345X”

‘ Enter Calibration Mode

‘ Apply precision voltage

PRINT#1, “OUTPUT07;H545X”

‘ Will calculate offsets for slave card 17 (TempScan/1100)

‘ Wait for command to complete

‘ Apply precision voltage of -50.0 mV

PRINT#1, “OUTPUT07;G545,1X”

‘ Calibrate negative gain for thermocouple, for slave card 17

(TempScan/1100)

‘ Wait for command to complete

‘ Apply precision voltage of +50.0 mV

PRINT#1, “OUTPUT07;G545,1X”

‘ Calibrate positive gain for thermocouple, for slave card 17

(TempScan/1100)

‘ Wait for command to complete

‘ Apply temperature of 90.0 °C

PRINT#1, “OUTPUT07;J21,2,90.0X”

‘ Calibrate cold junction offset for Channel 21 on the master

unit card

‘ Wait for command to complete

‘ Remove precision voltage

PRINT#1, “OUTPUT07;EX”

‘ End Calibration Mode

This manual is related to the following products: