Clear channel configuration – Measurement Computing TempScan/1100 User Manual
Page 210

A-14 API Command Reference
889897
TempScan / MultiScan User’s Manual
*C
- Clear Channel Configuration
TYPE
Channel
EXECUTION
Deferred
SYNTAX
*C
Clear out the channel configuration so that no channels are configured
DESCRIPTION
The Clear Channel Configuration (
*C
) command will clear out the channel configuration so that no channels are
configured. After this command has been performed, channel readings will be not be updated; therefore, no HLL
or Acquisition Buffer updates will be performed until a new configuration is entered via the Configure Channels (
C
)
command. This
*C
command may be used when the present configuration is no longer acceptable and you wish
to reconfigure channels from scratch.
EXAMPLE
PRINT#1,"OUTPUT07; *CX"
‘ Clear all channel configurations
PRINT#1,"OUTPUT07; C1,1X"
‘ Configure Channel 1 of thermocouple Type J