TempScan / MultiScan User’s Manual

889897

API

Command Reference A-29

D#

- Set Relay Make Time

MultiScan/1200 only

TYPE

System

EXECUTION

Deferred

SYNTAX

D#make

Specify how long to wait after closing a channel relay before reading valid

data, where

make

is the number of 520.833-microsecond intervals.

DESCRIPTION

The Set Relay Make Time (

D#

) command allows the user to increase the delay between closing a relay and

reading valid channel data. The system powerup default is:

make

=

6

= 3.125 milliseconds.

If common mode voltages on adjacent channels are widely dissimilar, reading inaccuracies can result. This
inaccuracy is due to inadequate settling time at the instrumentation amplifier when the unit is scanning between
channels. To eliminate these inaccuracies, the settling time can be increased using the

D#

command as shown

in the example below.

EXAMPLE

PRINT#1, “OUTPUT07;M#0X”

‘ Select line cycle integration/high-speed multichannel mode

PRINT#1, “OUTPUT07;W#32X”

‘ Select number of samples = 32

PRINT#1, “OUTPUT07;D#12X”

‘ Set Relay Make Time = 12 = 6.25 milliseconds

PRINT#1, “OUTPUT07;Y0,10,0X”

‘ Setup to read 10 past trigger scans

PRINT#1, “OUTPUT07;C1-360,1X

‘ Enable Channels 1 - 360 for Type J thermocouples

PRINT#1, “OUTPUT07;T1,8,0,0X”

‘ Arm the unit

PRINT#1, “OUTPUT07;@X”

‘ Trigger unit and collect scans