6-12 System Calibration

879596

TempScan / MultiScan User's Manual

Calibration of Voltage Scanning Cards

Note: Prior to each calibration session, disable calibration protection by setting the rear panel

DIP microswitch 9 to the up (1) position. Then enable Calibration Mode via the Enter
Calibration Mode
(

K

) command with a valid password. If the default calibration

password has not been changed, enter

K12345

.

Note: At the end of each calibration session, disable Calibration Mode with the End Calibration

Mode (

E

) command. Then re-enable calibration protection by resetting DIP

microswitch 9 to the down (0) position.

To Calibrate a Voltage Card for Offset

1. Connect a Copper Short to Channel 1 on the scanning card being calibrated. Use standard copper

hookup wire. Do not use thermocouple wire. Slide the scanning card into the previously-calibrated
master chassis.

2. Power up the unit and wait at least one hour before continuing.

3. Send the Calibrate Channel Offset (

H

) command

Hchan

– where

chan

refers to the first channel

number on the scanning card being calibrated (refer to the previous table for valid offset channel
numbers) – and wait for the command to complete. The TRIGGER LED indicator will flash for a few
seconds.

4. Monitor the Calibration Status Register (

CSR

) to ensure that the Calibrate Channel Offset (

H

)

command completed without error. Refer to the User Status (

U

) command

U2

to monitor and decode

the Calibration Status Register (

CSR

).

To Calibrate a Voltage Card for Gain

CAUTION

Ensure that the voltage scanning card has been calibrated for offset. If not, follow
the volts card offset calibration procedure as described in the previous text.

1. Connect a precision voltage source to Channel 1 on the scanning card being calibrated. Use standard

copper hookup wire. Do not use thermocouple wire. Slide the scanning card into the previously-
calibrated master chassis.

2. Power up the unit and wait at least one hour before continuing.

3. Set the Part A precision voltage source and send its corresponding Calibrate Channel Gain (

G

)

command

Gchan,type

– where

chan

refers to the first channel number on the scanning card being

calibrated (refer to the previous table for valid gain channel numbers) – from the following list:

Part A: -100.0 mV DC

Gchan,11

Part E:

-5.00 V DC

Gchan,13

Part B: +100.0 mV DC

Gchan,11

Part F:

+5.00 V DC

Gchan,13

Part C: -1.00 V DC

Gchan,12

Part G:

-10.00 V DC

Gchan,14

Part D: +1.00 V DC

Gchan,12

Part H:

+10.00 V DC

Gchan,14

Wait for the Part A command to complete. The TRIGGER LED indicator will flash for a few seconds.

4. Monitor the Calibration Status Register (

CSR

) to ensure that the Calibrate Channel Gain (

G

) command

completed without error. Refer to the User Status (

U

) command

U2

to monitor and decode the

Calibration Status Register (

CSR

).

5. Repeat Steps 3 and 4 for each of the next seven parts: Part B through Part H.