Data list, Data list -24 – Yokogawa Data Acquisition with PID Control CX2000 User Manual
Page 216

7-24
IM 04L31A01-17E
Data List
• The MI/MO command is used to output the data.
• The number of blocks and file names of the specified data in the internal memory are
output.
• If the first parameter of the MI/MO command is DIR, the data in the internal memory is
put on standby and the list is output.
Syntax
EA
CRLF
aa
CRLF
bb_ffffffff_eee_ssssss_yy/mo/dd_hh:mi:sskllllllllll
CRLF
.......................................................
EN
CRLF
aa
Number of valid blocks (00 to 99)
bb
Block number (00 to 99)
ffffffff
File name (8 characters)
eee
Extension (3 characters)
ssssss
Number of collections (_____1 to 999999)
yy
Year (00 to 99)
mo
Month (01 to 12)
dd
Day (01 to 31)
hh
Hour (00 to 23)
mi
Minute (00 to 59)
ss
Second (00 to 59)
The year/month/day/hour/minute/second will contain the time when
the last data of the block were acquired.
k
Data attributes
* Internal memory block being sampled
+ Internal memory block being overwritten
Space
Fixed block
If the data of the original block are changing when the data is put on
standby, * or + is output.
llllllllll
ID number (_________0 to 9999999999)
• The output is numerical value only when the file extension is CEV
or CDS. This value is specific to the data.
• The output is a space when the file extension is not CEV or CDS.
_
Space
Example
EA
02
01 DATA0001 DHR 128 99/02/24 20:10:00
02 DATA0002 DDR 128 99/02/24 20:11:00
EN
7.2 Output Format of ASCII Data