Yv sets the e-mail recipient address, Yw sets the e-mail sender address, Yx sets the e-mail smtp server name – Yokogawa Data Acquisition with PID Control CX2000 User Manual
Page 177

6-48
IM 04L31A01-17E
Example
Send report generation notification e-mail
messages including the source URL to recipient
1. The subject is “Report”, and the header is
“LP2”.
YUREPORT,ON,OFF,ON,Report,LP2
Description • For the contents of the system notification,
see section 2.11.
• Report generation notification can be used on
models with the computation function option /
M1.
• For details on the settings of e-mail, see
sections 2.3 and 2.11.
YV
Sets the e-mail recipient address
Syntax
YV p1,p2
p1: Select Recipient
1: Recipient 1
2: Recipient 2
p2: Recipient address (up to 150 characters)
Query
YV[p1]?
Example
Set recipient 1 to “[email protected]” and
YV1,[email protected] [email protected]
Description • To set multiple recipients, separate each
recipient with a space.
• For details on the settings of e-mail, see
sections 2.3 and 2.11.
YW
Sets the e-mail sender address
Syntax
YW p1
p1: Sender address (up to 64 characters)
Query
YW?
Example
Set the sender address to “CX2000”.
YWCX2000
Description For details on the settings of e-mail, see sections
2.3 and 2.11.
YX
Sets the e-mail SMTP server
name
Syntax
YX p1,p2
p1: SMTP server name (up to 64 characters)
p2: Port number (0 to 65535).
Query
YX?
Example
Set the SMTP server to “mhs.good.com” and
port number to “25”.
YXmhs.good.com,25
Description For details on the settings of e-mail, see sections
2.3 and 2.11.
6.7 Basic Setting Commands (Measurement)
ML
Sets the communication buffer
recovery processing
Syntax
ML p1
p1: DO/internal switches processing
Continue:
hold the status
Clear:
clear the status
Query
ML?
Example
During communication buffer recovery, clears
DO and the internal switches.
MLclear
MM
Sets the modbus mouse/
temperature meter
communication automatic
recovery
Syntax
MM p1,p2
p1:
Turns automatic recovery On/Off
p2:
Intervals: 1min, 2min, 5min, 10min, 20min,
30min, or 60min.
Query
MM?
Example
Temperature meter communications
automatically restored after 5 minutes.
MMON,5min