beautypg.com

Recording mode, Data length – Yokogawa Touch Screen GP20 User Manual

Page 94

background image

1-80

IM 04L51B01-01EN

Recording mode

Set the mode for recording event data.

Options

Description

Free

Records data at all times

Single

Records data when the trigger condition is met

Repeat

Records data every time the trigger condition is met

• Free
Recording starts when you start it and stops when you stop it.

Time

File

File

File

Adding data

• Single
The GX/GP enters the trigger-wait state when you start recording. After a trigger event

occurs, the GX/GP will record data for the specified time (data length) and stop. From this

point, the GX/GP will not record even if the trigger condition is met.

Time

File

Trigger condition met

• Repeat
The GX/GP enters the trigger-wait state when you start recording. After a trigger event

occurs, the GX/GP will record data for the specified time (data length) and stop. Then,

it enters the trigger-wait state again and repeats recording for the specified time (data

length) every time the trigger condition is met. To stop recording event data, stop the

recording.

Time

File

File

Trigger condition met Trigger condition met

Data length

Set the size of recording data per file. Recording data is divided at the specified file size.

The selectable data lengths vary depending on the number of channels to record and the

Recording Interval setting.
A file is also created in the following instances.

• When a file is created manually

• When recording is stopped

• When file creation is executed with the event action function

• After recovering from a power failure

Recording Interval

1

100 ms

200 ms

500 ms

1 s

2 s

Selectable range of data length 10 min to

1 days

10 min to

2 days

10 min to

3 days

10 min to

7 days

10 min to

14 days

Recording Interval

1

5 s

10 s

15 s

20 s

30 s

Selectable range of data length 10 min to

31 days

10 min to

31 days

10 min to

31 days

10 min to

31 days

1 hour to

31 days

Recording Interval

1

1 min

2 min

5 min

10 min

15 min

Selectable range of data length 1 hour to

31 days

1 hour to

31 days

1 hour to

31 days

1 hour to

31 days

1 hour to

31 days

Recording Interval

1

20 min

30 min

Selectable range of data length 1 hour to

31 days

1 hour to

31 days

1 You cannot choose a recording interval that is shorter than the scan interval.

1.8 Setting Recording Conditions (Recording mode, recording interval, saving interval)

This manual is related to the following products: