201
Cannot execute because other users changed setting.
Perform the operation again.
202
This action is not possible because recording or math computation is
in progress.
Execute after stopping recording or computing.
203
This action is not possible because sampling is in progress.
Execute after stopping recording.
204
This action is not possible because math computation is in progress. Execute after stopping computing.
211
This function is not possible at this time.
Check the GX/GP status, and then execute again.
221
This function is not possible because input number is over.
The limit is 50 messages (10 for free messages).
222
The message is not written while recording is stopped.
Execute after starting recording.
223
It is out of range for the postscript message.
There are no data files that messages can be written
in.
224
Message writing is not made outside a data range.
Move within the data range.
225
Failed to print.
For snapshot printing, wait a moment, and re-
execute.
Check the printer and network status.
226
Cannot perform action because message being written.
Wait until the additional message writing is finished.
231
This action is not possible because saving is in progress.
Wait until saving is complete.
232
This action is not possible because formatting is in progress.
Wait until formatting is complete.
233
Data save is not possible because of insufficient media capacity.
Replace the external storage medium.
235
This action is not possible because FTP transmission is in progress. Execute after FTP data transfer is complete.
236
Cannot take successive snapshots. Wait 10 seconds or more in
between snapshots.
Wait 10 seconds or more
251
The user name or the password is wrong.
Enter the correct name or password.
254
This entry is incorrect.
The maximum number of login users has been
exceeded.
255
Password entered is incorrect.
Enter the correct password.
256
You cannot use the same password.
Specify a different password.
257
Password is incorrect.
Enter the correct password.
258
This function is locked.
Log in using a user account that has permission to
use the operation.
260
This user name is unable to use this mode.
Users cannot switch modes.
281
Exceeds time deviation setting.
The time change made during recording exceeds
the time deviation limit. Set the time within the Time
deviation limit, which is set in Time basic settings of
System settings. page 1-143 in section 1.18.4, “Setting
the Time Zone, Gradual Time Adjustment, and
• Operation Errors
Code
Message
Description, Corrective Action, Ref. Section
501
Operation aborted because an error was found on media.
Change the storage medium.
502
Not enough free space on media.
Not enough free space on the storage medium, or the
limit of the number of directories has been exceeded.
Change the storage medium.
503
Media is not recognized.
Remove the storage medium, and set it again.
504
Media read error.
Change the storage medium.
505
Media write error.
506
Failed to create the file.
507
There is no file or directory.
Tried to access a file which does not exists in the
internal memory.
508
Format error.
Reformat.
509
Unknown file type.
Check the extension.
511
Invalid file or directory operation.
Tried to delete a directory containing files. Delete
the files and directories in the directory first before
executing the operation.
512
The file is already in use. Try again later.
Wait until the file is accessible.
514
There is no setting file.
Check the setting parameter file.
515
A template file is abnormal.
Check the report template file.
516
There are items which was not set.
Check the error log.
517
Media is read-only.
Make it writable.
518
Scale image not loaded.
Load a scale image.
519
Template file not loaded.
Load a template file.
532
There is no available data.
This error occurs when there is no valid data in the
internal memory. Specify valid data.
536
There is no data after the selected time for this day.
Select an earlier time.
537
The specified screen cannot be displayed
Specify a valid screen.