beautypg.com

Explanation, High limit alarm and low limit alarm, Delay high limit alarm and delay low limit alarm – Yokogawa Touch Screen GP20 User Manual

Page 34

background image

1-20

IM 04L51B01-01EN

Explanation

Alarm Type
The character inside the parentheses is the symbol denoting each alarm.

High Limit Alarm (H)

Alarm value

Alarm output ON

Alarm output ON

Hysteresis

Measured value

Low Limit Alarm (L)

Alarm output ON

Alarm value

Measured value

Hysteresis

Alarm output ON

Delay period

Alarm value

Measured value

Delay High Limit Alarm (T)

Delay Low Limit Alarm (t)

Alarm output ON

Delay period

Alarm value

Measured value

High Limit on Rate-of-Change Alarm (R)

Low Limit on Rate-of-Change Alarm (r)

Alarm value

Hysteresis

Difference in the measured values of two channels

Difference High Limit Alarm (h)

Difference Low Limit Alarm (l)

Alarm output ON

Alarm value

Hysteresis

Difference in the measured values of two channels

Change in the

measured value

Measured value

Time

Interval (t2 – t1)

T2

T1

t1

t2

Change in the measured value

Measured value

Time

Interval (t2 – t1)

Amount of change

in the setting

|

|

T2

T1

t1

t2

T1 – T2

Amount of change

in the setting

|

|

T1 – T2

High Limit Alarm and Low Limit Alarm

An alarm is activated when the measured value is greater or less than or equal to the alarm

value.

Delay High Limit Alarm and Delay Low Limit Alarm

An alarm is activated if measured values remain greater or less than or equal to the alarm

value for a specified time period (delay period).

High Limit on Rate-of-Change Alarm and Low Limit on Rate-of-Change Alarm

An alarm is activated if the increasing or decreasing rate-of-change of measured values over

a certain interval is greater than or equal to the specified value.

The alarm value of an rate-of-change alarm is set using an absolute value. The interval is

derived using the following equation and set using the number of samples.

Interval = the scan interval × the number of samples

For the number of samples, see page 1-142 in section 1.18.2, “Setting the Interval for

Calculating the Rate-of-Change for Rate-of-Change Alarms”.

Difference High Limit Alarm and Difference Low Limit Alarm

An alarm is activated when the difference in the measured values of two channels is greater

or less than or equal to the specified value.

This alarm can be specified on measurement channels set to difference calculation.

1.2 Configuring AI Channels (Analog input (including DI) channels)

This manual is related to the following products: