2-2
IM 04L51B01-01EN
Stopping Recording
For the procedure when the batch function is in use, see page 2-3 in section 2.1.2, “Using the
1
Press
MENU.
The menu screen appears.
2
Tap
Record.
The record stop screen appears.
3
Tap
Rec+Math or Record.
Recording stops. The recording status icon indicates stopped.
When computation (/MT option) is in progress, you can select
Rec+Math. Computation
stops when recording stops.
Operation complete
• You can also use the
START/STOP key. If you use this key, recording will stop
immediately.
• If
START/STOP key action in the math channel settings is set to Start / Stop or Reset +
Start / Stop, computation also stops when you press the START/STOP key.
Explanation
Operations That Start When Recording Starts
• Waveform display updating on the trend display
• Reporting (/MT option)
• Computation (/MT option; when
START/STOP key action is set to Start / Stop or Reset
+ Start / Stop)
Operations That Stop When Recording Stops
• Waveform display updating on the trend display
• Reporting (/MT option)
• Computation (/MT option; depends on the setting specified in the procedure above)
Performance While Data Is Being Saved
If the internal memory or external storage medium is continuously accessed, the following
phenomena may occur. When such phenomena occur, the storage medium access indicator
frequently illuminates.
• Files being saved to the external storage medium drop out.
• Accessing the GX/GP through communication takes a long time.
If these conditions occur, take the following measures.
• If you are creating data files at short intervals consecutively using the event action
function, increase the data file save interval.
• If you are creating numerous files in a single directory on the external storage medium,
change the destination directory name at approximately every 1000 files.
• If data recording and display are using up resources (for example if you are recording
on multiple channels at a fast rate and displaying trend displays on all the multi panels),
reduce the recording interval, or change the display.