1.3 Configuring DI Channels (Digital input channels)
Set the DI channels’ input range, alarm, and display conditions.
To use the DI module as a remote control input, set the operation mode of the module to
Remote. Refer to page 1-76 in section 1.7.3, “Setting the Operation Mode of a Module”.
Set the remote control action using event action. Refer to page 1-108 in section 1.14,
“Configuring the Event Action Function”.
1.3.1
Setting the Range
Path
GX/GP:
MENU key > Browse tab > Setting > Setting menu DI channel settings > Range
Web browser:
Config. tab > DI channel settings > Channel range (display example: 0501-
0516)
Hardware configurator:
DI channel settings > Channel range (display example: 0501-0516)
Description
Setup Item
Selectable Range or Options
Default Value
First-CH
DI channel
—
Last-CH
DI channel
—
First-CH, Last-CH
Set the target channels. The channels that you can specify appear depending on the module
configuration.
Range
Setup Item
Selectable Range or Options
Default Value
Type
Skip, DI
DI
Span Lower
0, 1
Delta: –1 to 1
0
Span Upper
0, 1
Delta: –1 to 1
1
Calculation
Off, Delta, Linear scaling
Off
Reference channel
1
Value (I/O channel)
—
1 Appears when Calculation is set to
Delta.
Type
Set the input type.
Options
Description
Skip
Not measured.
DI
Displays contact input or voltage input signals by mapping them to 0% or 100% of the
display range.
Span Lower, Span Upper
Set the input range.
Calculation
When performing input calculation, set the calculation type. You can set this when the range
type is set to
DI.
For details on the input calculation function, see page 1-11 in section 1.2.1, “Setting the
Reference channel
Set the reference channel for the difference calculation.