ALLIED Vision Technologies Stingray F-504 User Manual
Page 198

Controlling image capture
STINGRAY Technical Manual V4.4.4
196
Note
•
Exposure times entered via the 81Ch register are mir-
rored in the extended register, but not vice versa.
•
Longer integration times not only increase sensitivity,
but may also increase some unwanted effects such as
noise and pixel-to-pixel non-uniformity. Depending on
the application, these effects may limit the longest
usable integration time.
•
Changes in this register have immediate effect, even
when the camera is transmitting.
•
Extended shutter becomes inactive after writing to a
format/mode/frame rate register.