Appendix D: Performance Verification

MTG100 & MTG300 MPEG Generator User Manual

DĆ17

MPEG generator

rear panel

75 W BNC cable

Signal generator (HP 8648)

50 W BNC cable

NĆtoĆBNC adapter

MPEG test systemrear panel

(MTS200 or MTS215 Opt. SS)

Figure DĆ10: Equipment connections for verifying the external reference input

5. Select Clock... from the OUTPUT menu on the MPEG generator to open

the Clock dialog box.

6. Select EXT REF in the Clock dialog box, and then press the OK bezel

button.

7. Press the START/STOP button on the MPEG generator to start the signal

output of the test64.trp file.

8. Verify that the following LEDs on the MPEG generator’s front panel are

illuminated: PLL, BUFFER, EXT, and HDD.

9. Double-click the Real Time Analyzer icon in the MPEG test system’s

Program Group window to start the application.

10. In the Real Time Analyzer window on the MPEG test system, select

Settings... from the Configuration menu to open the Settings window.

11. In the Settings window, select Hardware Configuration to open the

Hardware Configuration window.

12. In the Hardware Configuration window, click the Restore Defaults button,

click the Apply button, and then select DVB-PI ASI as the Input Type.

13. In the Settings window, select ETR290 to open the ETR290 configuration

window.

14. In the ETR290 window, click the Restore Defaults button, click the Apply

button, and then click OK to close the Settings window.