Appendix D: Performance Verification

DĆ12

MTG100 & MTG300 MPEG Generator User Manual

This test confirms that transport stream data is output correctly in serial format
from the ECL/TTL/LVDS OUT connector on the MPEG generator. The
following equipment and MPEG test signal is required for this test:

H MPEG test system

H Parallel interface cable

H test64.trp MPEG test signal

Perform the following procedure to verify that transport stream data in parallel
format is output correctly from the ECL/TTL/LVDS OUT connector on the
MPEG generator:

1. Use the parallel interface cable to connect the ECL/TTL/LVDS OUT

connector on the MPEG generator to the ECL PARALLEL/SERIAL I/O
connector on the MPEG test system rear panel as shown in Figure D-8.

2. Open the test64.trp file on the MPEG generator.

a. Select Open TS File... from the FILE pull-down menu.

b. In the resulting Select TS File dialog box, select the test64.trp file, and

then press the OK bezel button.

3. Select Level : ECL from the OUTPUT menu on the MPEG generator, and

then press the OK bezel button.

4. Select Interface : Serial from the OUTPUT menu on the MPEG generator,

and then press the OK bezel button.

5. Press the START/STOP button on the MPEG generator to start the signal

output of the test64.trp file.

6. Verify that the following LEDs on the MPEG generator’s front panel are

illuminated: PLL, BUFFER, and HDD.

MPEG generator rear panel

Parallel interface cable

MPEG test system rear panel

(MTS200 or MTS215 Opt. SS)

Figure DĆ8: Equipment connection for verifying the serial output

Serial Output