Calibrate maximum level:

U

Press the MAX soft key

U

Use the arrow cursor keys to select the same channel as in step 1
if necessary

U

Position the channel probe against the measurement Max reference
surface for the maximum end of the channel calibration range

U

Use the numeric keypad to enter the Max value into the Master
Value data field for the channel

U

Press the ENTER key

U

Press the YES soft key to calibrate the maximum value for the
channel input

The completed maximum level calibration will be indicated by a filled
green circle in the Mx column at the right of the Actual channel value.

Completed maximum level calibration.