Locking the calibration process if warnings occur, Specifying a calibration interval, 3 s o ft wa re s e tu p – HEIDENHAIN ND 2100G User Manual
Page 112

112
2 Installation and Specifications
2
.3
S
o
ft
wa
re
s
e
tu
p
Locking the calibration process if warnings occur
The calibration process can automatically be locked when a calibration
fails and a warning is displayed. Once locked, the calibration process
can only be completed by performing a new calibration that is
successful or by entering the supervisor password to abort the
calibration process.
To lock calibration when a warning is displayed:
U
Highlight the Lock If Warning field
U
Press the YES soft key
U
Press the ENTER key
Specifying a calibration interval
The Master calibration screen can automatically be displayed at
regular intervals to support a schedule of periodic input device
calibrations. The Master screen will be displayed at the specified
intervals unless a setup screen is being displayed to configure the
ND 2100G.
To specify a calibration interval:
U
Highlight the Do Every (hrs) field
U
Use the numeric keypad to enter the amount of time in hours
U
Press the ENTER key
The user will be unable to skip a calibration without the
supervisor password when the Lock If Warning field is set
to Yes.