beautypg.com

Locking the calibration process if warnings occur, Specifying a calibration interval, 3 s o ft wa re s e tu p – HEIDENHAIN ND 2100G User Manual

Page 112

background image

112

2 Installation and Specifications

2

.3

S

o

ft

wa

re

s

e

tu

p

Locking the calibration process if warnings occur

The calibration process can automatically be locked when a calibration
fails and a warning is displayed. Once locked, the calibration process
can only be completed by performing a new calibration that is
successful or by entering the supervisor password to abort the
calibration process.

To lock calibration when a warning is displayed:

U

Highlight the Lock If Warning field

U

Press the YES soft key

U

Press the ENTER key

Specifying a calibration interval

The Master calibration screen can automatically be displayed at
regular intervals to support a schedule of periodic input device
calibrations. The Master screen will be displayed at the specified
intervals unless a setup screen is being displayed to configure the
ND 2100G.

To specify a calibration interval:

U

Highlight the Do Every (hrs) field

U

Use the numeric keypad to enter the amount of time in hours

U

Press the ENTER key

The user will be unable to skip a calibration without the
supervisor password when the Lock If Warning field is set
to Yes.