SLEC setup procedure
The SLEC setup procedure includes the following steps:
Clear all datums
The following steps clear datums in preparation for SLEC standard
measurements.
To clear all datums:
From the DRO screen.
U
Press the MENU... soft key
U
Press the DATUM... soft key
U
Press the CLEAR soft key
U
Press the YES soft key to confirm clearing all datums
Clear all
Datums
Cycle Power &
Set Machine 0
Clear Old
Values &
Disable SLEC
Enter
Machine 0
Offset
Enter New
Station
Values
Enable SLEC
Make sure that all encoder channel resolutions and
machine zero references are properly configured, and that
any required channel Master operations have been
completed prior to configuring the SLEC feature. Refer to
the description of Channel setup on page 99, and to the
description of the Master function in Chapter 1: Operation,
on page 35.
Clear all
Datums
Cycle Power &
Set Machine 0
Clear Old
Values &
Disable SLEC
Enter
Machine 0
Offset
Enter New
Station
Values
Enable SLEC