Calibration of a single reference point

A single measurement reference point can be calibrated for encoders
and fully-calibrated transducers.

To calibrate a single reference point

U

Press the MASTER soft key

U

Use the arrow cursor keys to select the desired channel

U

Position the probe against the reference surface

U

Use the numeric keypad to enter the reference value (zero or the
desired offset) into the Master Value data field

U

Press the ENTER key

U

Press the YES soft key to calibrate the measurement reference

A green dot will appear next to the actual value indicating that the
reference point is calibrated.

Channel 1 selected. Reference value of 0.

Press the YES soft key.

Channel 1 calibrated.

The Master type should be set to Mean in the Master
setup screen to perform a calibration of a single encoder
point. Refer to "Master setup screen" on page 111.