Switch view, Cpk/ppk display, Slew limit – HEIDENHAIN ND 2100G User Manual
Page 159: Start scr saver, 3 s o ft wa re s e tu p

ND 2100G GAGE-CHEK
159
2
.3
S
o
ft
wa
re
s
e
tu
p
Switch view
Normally, the user presses soft keys to select the screens displayed
on the LCD. However, the ND 2100G can be programmed to
automatically change the DRO, bar or dial views to display a dimension
in response to activity on the channel included in its formula. The
Switch View field is used to specify the range of motion required on
any channel input to switch DRO, bar or dial views to the associated
dimension. When the Switch View field is zero, no switching occurs.
To specify Switch View:
U
Highlight the Switch View field
U
Use the numeric keypad to enter the range of motion required to
switch the DRO view
U
Press the ENTER key
Cpk/Ppk display
Cpk/Ppk Display is used to select process capability or process
performance data to be displayed in the Data view screen for a single
dimension.
To specify Cpk Cp or Ppk Pp data to be displayed:
U
Highlight the Cpk/Ppk Display field
U
Press the PPK PP or CPK CP soft key to select process or capability
data
U
Press the ENTER key
Slew limit
High input slew rates resulting from rapid input encoder motion can
result in erroneous measurements. Erroneous measurements are
avoided by displaying encoder error warnings when encoder values
change at very high rates.
Enter the slew rate limit into the Slew Limit field in counts (increments
of resolution) per second. For example, at a channel resolution of
0.001 mm, a slew rate limit of 50,000 will result in warning messages
at encoder motion rates higher than 50 mm per second.
To specify Slew Limit:
U
Highlight the Slew Limit field
U
Use the numeric keypad to enter the slew limit
U
Press the ENTER key
Start Scr saver
The Start Scr Saver field is used to specify the minutes of ND 2100G
inactivity before the LCD screen saver is displayed. The screensaver
is disabled when 9999 is entered into this field.
To specify Start Scr Saver:
U
Highlight the Start Scr Save field
U
Use the numeric keypad to enter the number of minutes of inactivity
before the screen saver is activated
U
Press the ENTER key