27 calib. settings, Calib. settings -30 – Yokogawa GC8000 Process Gas Chromatograph User Manual
Page 77

<4. Display and Operations while Connected to GC8000>
4-30
IM 11B06C01-01E
4.3.27
Calib. Settings
To set various types of data related to calibration, select the Calib. Settings tab window from the
Set Peak dialog box.
Figure 4.34
Set Peak dialog box (Information Related to Calibration)
In this dialog box, you can set the Standard Concentration, Standard Area, and Calibration
Coefficient for the specified stream number, system number, and peak number. Note that the
information about a stream not being set as the Calibration Stream is grayed out and you cannot
change it.
After you have changed the settings, click the Change button. A confirmation message asking
“Are you sure you want to send the data?” is displayed in the message box. When you respond
with “Yes”, the settings are changed.
5th Edition : Nov. 16, 2011-00